Fume cupboards - Flipbook - Page 9
Product specification
Ergo Zafe Pathology 19
Fume cupboard
Model
1550
1850
2150
2450
Ergo zafe Patologi 19
3050
Designed in accordance with EN14175. Certificate available on request.
Colour enclosure and ext.
sides
White
Width
1550 mm
Depth
1017 mm
Height
2500 mm
Int. enclosure height
890 mm
Height of worktop
720-1020 mm
Working depth of worktop
699 mm
Sash opening
180-550 mm
Upper limit fume cupboard
Max 2500 mm
Front sash
Acrylic
Light fixture
HF light 940, 840 or 830.
Or LED light Ra 90k4000, Ra90k3000, Ra80k3000
Ergo
Dimming option
1850 mm
2150 mm
2450 mm
3050 mm
Ergo z
Kan le
valgfri
Låge m
Forber
opstilli
Lugeå
Stinkskab
højde 2500mm.
263
A
zafe Patologi
Option19
Hoisting system
Motorised
Chamber liner
High-pressure laminate
Designed in accordance with ISO 3746
Ventilation connection (ILU) 2xØ200
mm
1581
1851
2121
2661
2xØ200
mm
2xØ200
mm
2xØ200
mm
2xØ200
mm
Ergo zafe Patologi 19
Pressure drop in cabin
Approx. 9 Pa at 400m3/h
Water and sanitation
connections
½" union connection
Drainage connections
Kan leveres
med 1xdobbelt
70mm bottom valve with
1½" pipe
thread230V indvendig stikkontakt,
Lance for washing
connection
½" pipe thread
Electrical connections
3x1.5 PKAJ (L1 + N + PE) (See wiring diagram)
525
882
1,311
Stinkskab højde 2500mm.
126
A
1017
Y:TegningerBirgitteNyt katalog 2014Bilag 104.idw
Ergo zafe Patologi 19, Højde 2500mm
valgfri placering til højre eller venstre.
Låge med motorhejs.
Forberedt for monkeybars, der giver mulighed for høje
opstillinger eller hylder.
Lugeåbning: 170-562mm
Ergo zafe Patologi 19, Højde 2500mm.
Kan leveres med 1xdobbelt 230V indvendig stikkontakt,
valgfri placering til højre eller venstre.
Låge med motorhejs.
Forberedt for monkeybars, der giver mulighed for høje
opstillinger eller hylder.
Lugeåbning: 170-562mm.
Ventilation connection base Ø63 mm extract spigot
cabinet
Ventilation connection sink
Ø84 mm extract spigot
Actuator capacity
3200 N
Max weight load A
capacity on base cabinets
and worktop
Min. 150 kg.
per rm
Operational temperature
5 to 40 deg.
3200 N
6400 N
6400 N
6400 N
B
263
A
Ø200
Approx. 2 metre cable for electrical
connection
B
263
Ø200
B
1017
Y:TegningerBirgitteNyt katalog 2014Bilag 104.idw
Zystm A/S • Lillelundvej 18-20126• DK 7400 Herning
• Tel.:
9666 6660 • info@zystm.com
• www.zystm.com
B
BACK TO
CONTENT
A
Y:TegningerBirgitteNyt katalog 2014Bilag 104.idw
1017
1690-1990
720-1020mm
126
126
Min. 700mm
780-1080mm.
882
A
2500
882
2500
525
525
1690-1990
Max. Volume flow m /h
sash opening of 500 mm
and 0.5 m/s
3
Min. 700mm.
Noise
Ø200
126